AccreditationWhy emergency power

Meeting Standards and Requirements for the VFC, CDC and NSF456

What Is Vaccine Storage Standard NSF 456?

The NSF 456 standard for vaccine storage was developed collaboratively by the CDC and NSF International. Though not legally mandated, this standard is instrumental in ensuring vaccines and medications are maintained at proper temperatures, helping safeguard their effectiveness.

While healthcare facilities aren’t required to purchase new refrigerators or freezers, selecting those that comply with NSF 456 can significantly cut costs by reducing the loss of vaccine doses.

Temperature Thresholds for Testing

During validation testing for compliance:

Refrigerators

  • During door-opening tests, the temperature probe must remain between 2 °C (35.6 °F) and 8 °C (46.4 °F).
  • After a three-minute door-open test, the unit must recover to 5 °C (41 °F) ± 3 °C within that recovery period.

Freezers

  • During door-opening tests, the probe must stay between −50 °C (−58 °F) and −15 °C (5 °F).
  • After one minute of door opening, the probe’s temperature must not rise above −15 °C (5 °F).

The Role of Backup Power

Although battery backup isn’t mandatory, having one is highly advantageous during power interruptions. It helps the unit continue operating at safe temperatures, protecting your vaccines, medications, and other sensitive items. Investing in a backup power system is a practical and cost-effective alternative to replacing spoiled biomedical stock.

Share this article

Further Reading

Case Study: Adapting a Two-Hour Backup Power Plan for an Operating Room in Virginia 
Surgery Centers

Case Study: Adapting a Two-Hour Backup Power Plan for an Operating Room in Virginia 

An operating room project in Virginia needed two hours of battery backup power for designated equipment. As construction progressed, the company identified additional equipment needing protection, and the backup power plan needed to reflect those requirements.  Medi-Products worked with the project team to review equipment specifications and electrical drawings, adjust the system sizing, and develop […]

Read more
Can One Battery Backup System Power Multiple Refrigerators and Freezers? 
Healthcare ComplianceRefrigerated Medications

Can One Battery Backup System Power Multiple Refrigerators and Freezers? 

A pharmacy may have two vaccine refrigerators and a freezer. A laboratory might need backup power for cold storage, incubators, and supporting equipment. Does every device need its own battery backup system?

Read more
Which Battery Backup System Will Work With My Vaccine Refrigerator, Pharmacy Fridge, or Laboratory Freezer? 
Refrigerated MedicationsVaccine Storage

Which Battery Backup System Will Work With My Vaccine Refrigerator, Pharmacy Fridge, or Laboratory Freezer? 

Choosing battery backup for cold storage starts with the equipment you need to protect. A vaccine refrigerator, pharmacy fridge, and laboratory freezer may have different electrical requirements, even when they look similar or occupy the same amount of space.  The right system must match the equipment’s electrical specifications, handle its startup demand, and provide enough […]

Read more